The TPS smartCHEM-D is a research-grade benchtop dissolved oxygen and conductivity/TDS analyser rated to IP65. Features a 240 × 64 dot graphic display, 1,489-reading datalogger, autolog, RS232 interface, Automatic Stability Function (ASF), automatic DO calibration in air or against zero-DO solution, salinity compensation, 4.5V DC stirrer output for the YSI self-stirring BOD electrode, PIN code access, and GLP calibration data storage. Designed and manufactured in Australia by TPS.
Science Essentials has supplied quality laboratory equipment and consumables to Australian laboratories since 1979. Brisbane-based with Australia-wide delivery.
Technical Information
Technical Information
Dissolved Oxygen
Ranges
ED1 :
0. 00 to 20. 00 ppM
20. 0 to 40. 0 ppM
0. 0 to 250. 0 % Sat'n
250 to 450 % Sat'n
0. 0 to 50. 0 % Gaseous
50 to 100 % Gaseous
YSI :
0. 00 to 25. 00 ppM
25. 0 to 40. 0 ppM
0. 0 to 300. 0 % Sat'n
300 to 450 % Sat'n
0. 0 to 60. 0 % Gaseous
60 to 100 % Gaseous
Resolution :
0. 01 1 ppM
0. 1 1 % Saturation
0. 1 1 % Gaseous
Accuracy :
+/-0. 2 % of full scale of selected ppM range
+/-0. 3 % Saturation
+/-0. 1 % Gaseous
Conductivity
TDS
Note
Dissolved Oxygen, Conductivity, and TDS ranges are automatically selected. Exact auto-range points are subject to sensor performance.
Specific Ions
Ranges : Auto-ranging in units of ppM, ppK, % and Exponential Notation
Resolution : User selectable for 3 significant digits, 2 significant digits or Auto-rounding
Accuracy : +/-Least significant digit
pH
Ranges :
0. 000 to 14. 000 pH
0. 00 to 14. 00 pH
Resolution :
0. 001 pH
0. 01 pH
Accuracy :
+/-0. 002 pH
+/-0. 01 pH
Millivolts Relative Millivolts
Range : 0 to +/-400. 0 and 0 to +/-1500 mV (auto-ranging)
Resolution :
0. 1 mV
1 mV
Accuracy :
+/-0. 15 mV
+/-1 mV.
Temperature
Range : -10. 0 to 120. 0 OC (sensor limit 45 OC)
Resolution : 0. 1 OC
Accuracy : +/-0. 2 OC
Temperature Compensation
0 to 100 OC, automatic or manual.
Calibration
Automatic calibration for all parameters. User can define standards where applicable.
Memory
1489 readings including date and time.
Automatic Datalogging
User-set for one reading every 2 to 90 seconds, minutes or hours.
RS232 Serial Port Output
1200, 9600, 19200 and 38400 Baud.
8 Bits, No Parity, 1 Stop Bit, XON/XOFF Protocol
Recorder Output (optional)
Max 0 to 2000mV output for any one parameter at one time.
Output impedance approx 1000 Ohms.
Resolution approx 2 mV.
Clock
24 Hour Calendar clock.
Displays date, month, year, hours, minutes, seconds.
Good Laboratory Practices
Date, time and value of last successful calibration for each parameter are stored. This information can be recalled or sent to the RS232 port at any time.
Calibration Alarm
Calibration Alarm can be activated for parameter of user's choice. Calibration Alarm can be set to Daily, Weekly, 2 Weekly or 4 Weekly.
Display
240 x 64 dot graphic LCD with large, bold digits, user-friendly menu system and full-text help and error messages.
Enclosure
Tough, high impact ABS. Splash resistant one-piece lid.
Dimensions
240 x 180 x 105 mm
Mass
Instrument only : Approx 1. 0 kg
Full Kit : Approx 5. 0 kg
Environment
Temperature : 5 to 45°C
Humidity : 0 to 90% R. H.
Power
12V DC via AC/DC adaptor.
AC/DC adaptor to suit specified country is supplied.
123104