The TPS smartCHEM-Ion3 is a research-grade benchtop specific ion/pH/mV analyser with three independent channels, each individually programmable for specific ions, pH, or mV. Rated to IP65. Features a 240 × 64 dot graphic display, 1,200-reading datalogger, autolog, RS232 interface, Automatic Stability Function (ASF), multi-point calibration with automatic buffer recognition, PIN code access, GLP calibration data storage with programmable calibration alarm, and Karl Fischer polarisation output. Designed and manufactured in Australia by TPS.
Science Essentials has supplied quality laboratory equipment and consumables to Australian laboratories since 1979. Brisbane-based with Australia-wide delivery.
Technical Information
Technical Information
Ranges : Auto-ranging in units of ppM, ppK, % and Exponential Notation
Resolution : User selectable for 3 significant digits, 2 significant digits or Auto-rounding
Accuracy : +/-Least significant digit
Ranges :
0. 000 to 14. 000 pH
0. 00 to 14. 00 pH
Resolution :
0. 001 pH
0. 01 pH
Accuracy :
+/-0. 002 pH
+/-0. 01 pH
Range : 0 to +/-400. 0 and 0 to +/-1500 mV (auto-ranging)
Resolution :
0. 1 mV
1 mV
Accuracy :
+/-0. 15 mV
+/-1 mV
Resolution : 0. 1 OC
Accuracy : +/-0. 2 OC
0 to 100 OC, automatic or manual.
Automatic calibration for all parameters.
User can define standards where applicable.
2730 readings including date and time.
User-set for one reading every 2 to 90 seconds, minutes or hours.
1200, 9600, 19200 and 38400 Baud.
8 Bits, No Parity, 1 Stop Bit, XON/XOFF Protocol
Max 0 to 2000mV output for any one parameter at one time. Output impedance approx 1000 Ohms.
Resolution approx 2 mV.
24 Hour Calendar clock.
Displays date, month, year, hours, minutes, seconds.
Date, time and value of last successful calibration for each parameter are stored. This information can be recalled or sent to the RS232 port at any time.
Calibration Alarm can be activated for parameter of user's choice. Calibration Alarm can be set to Daily, Weekly, 2 Weekly or 4 Weekly.
240 x 64 dot graphic LCD with large, bold digits, user-friendly menu system and full-text help and error messages.
Tough, high impact ABS. Splash resistant one-piece lid.
240 x 180 x 105 mm
Instrument only : Approx 0. 8 kg
Full Kit : Approx 3. 0 kg
Temperature : 5 to 45°C
Humidity : 0 to 90% R. H.
12V DC via AC/DC adaptor.
AC/DC adaptor to suit specified country is supplied.
121164